Left ventricular assist device driveline infections: recent advances and future goals
Review Article on Left Ventricular Assist Devices

Left ventricular assist device driveline infections: recent advances and future goals

Anne-Marie Leuck

Department of Medicine, University of Minnesota, Minneapolis, MN, USA

Correspondence to: Anne-Marie Leuck, MD. Department of Medicine, University of Minnesota, 420 Delaware Street. SE, Mayo Mail Code 250, Minneapolis, MN 55455, USA. Email: leuck002@umn.edu.

Abstract: Left ventricular assist devices (LVADs) have revolutionized the treatment of advanced heart failure, but infection remains a substantial risk. LVAD driveline infections (DLIs) are the most common type of LVAD-associated infection (LVADI). In the past several years we have expanded our understanding of DLI epidemiology, standardized the definition of LVADIs, improved infection rates through changes in implantation techniques, and investigated potential new modalities for DLI diagnosis. However, significant challenges remain for optimizing DLI prevention and treatment. These challenges include standardizing and improving both empiric and targeted antimicrobial therapy, expanding our understanding of effective driveline exit site dressings and topical therapies, and defining the patient population that benefits from device exchange and transplant. Additionally, in an era of expanding antibiotic resistance we need to continue investigating novel, non-antibiotic therapies for prevention and treatment of DLIs.

Keywords: Device-related infection; left ventricular assist devices (LVADs); driveline infections (DLIs); biofilm


Submitted Oct 19, 2015. Accepted for publication Oct 21, 2015.

doi: 10.3978/j.issn.2072-1439.2015.11.06


Introduction

Left ventricular assist devices (LVADs) have revolutionized the treatment of advanced heart failure, but infection remains a substantial risk. Although LVAD technology continues to advance rapidly, all current devices still require an external power source with energy supplied via a tunneled percutaneous driveline. Driveline infections (DLIs) are the most common type of LVAD-associated infection (LVADI). DLIs occur frequently because the driveline exit site creates a conduit for entry of bacteria and the prosthetic material of the driveline creates an ideal environment for the formation of bacterial biofilms (1). DLI, along with gastrointestinal bleeding (GIB) and stroke, are the leading causes of unplanned readmission for patients with LVADs, with a median direct hospital cost of more than $7,000 per readmission (2).

The prevalence of LVAD complications, including infection, is increasing as use of LVADs has expanded from short-term use as a bridge to transplantation (BTT) to also include long-term destination therapy (DT) in patients who are not eligible for transplant (3). Several recent reviews and retrospective studies have outlined the epidemiology and broad treatment approach for LVADIs (1,4-7). This review will focus on recent advances and persisting knowledge gaps related to diagnosis, specific treatments, and prevention of DLIs in adults.


Definitions and rates of infection

Definitions

In 2011, the International Society for Heart and Lung Transplantation (ISHLT) proposed consensus guidelines for the definition of LVADIs. Within the ISHLT definition, LVAD DLIs can be divided into deep and superficial infections. Both infections involve the soft tissue surrounding the driveline exit site and are accompanied by erythema, increase in temperature around the site, and purulent drainage. Deep infections also include the fascia and muscle layers (8). Since the exact extent of infection can often only be determined during surgical exploration, the distinction between superficial and deep infection is of limited utility in clinical care. Additionally, differentiating DLI from pump pocket infection, which is infection involving the body cavity that holds the LVAD pump, can also be difficult in the absence of surgical investigation.

Rates of infection

Incidence and prevalence of DLIs vary greatly between studies depending on the population evaluated and definitions used. In their 2011 review, Pereda and Conte note a range in DLI incidence of 14-48% in a sample of studies published from 2004 to 2010 (1). One difficulty in defining the epidemiology of DLIs has been the small sample size of most studies. The Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) allows for evaluation of larger patient cohorts involving multiple institutions. Using INTERMACS, Goldstein et al. evaluated 2,006 LVAD recipients who had a continuous-flow LVAD implanted from June 2006 to September 2010 and found a 19% prevalence of DLI 12 months after implantation (6).

More recent studies have attempted to go beyond incidence rates and evaluate risk factors for DLI, including age. Advanced age does not appear to be a risk factor for DLI with a recent study showing no difference in infection rates in patients over and under age 65 (9). However, younger age was actually found to be the only risk factor for DLI in a multivariate analysis by Goldstein et al. This finding is thought to be due to higher activity rates, and therefore increased risk for driveline exit site trauma, in the younger population (6). Trauma has previously been identified as a risk factor for subsequent DLI (10).


Diagnosis

Diagnosis of a DLI usually occurs when the patient, caregiver, or provider notes erythema, warmth, or purulent drainage around the driveline exit site. However, determining whether the infection is limited to the driveline exit site or involves deeper structures is difficult. There are no specific guidelines for the use of imaging to assess the extent of infection after diagnosis of a DLI. Ultrasound can detect fluid pockets but otherwise gives little information about whether structures are infected. Additionally, computed tomography (CT) is of limited utility given the artifact caused by the device (11).

Due to the limitations of standard imaging techniques, more advanced imaging options have been considered for evaluation of DLIs. One case series found that gallium single photon emission tomography-CT (SPECT-CT) may elucidate the extent of LVAD structures involved after diagnosis of DLI and help to inform decisions related to need for device exchange (11). However, it should be noted that previous publications have questioned the sensitivity of this imaging modality since inflamed but uninfected tissue might be misidentified as infection (3). Positron emission tomography (PET)-CT is also undergoing investigation as an imaging modality for LVADI and was useful in identifying infection of LVAD components and response to therapy (12,13). PET-CT may also reveal unsuspected distant sites of infection such as paravertebral abscess (12). Given the prolonged bacteremia that often accompanies DLIs, especially with pathogens such as Staphylococcus aureus or Pseudomonas aeruginosa, evaluation for metastatic sites of infection can be crucial.


Pathogens

The pathogens involved in cases of LVAD DLI are predominantly skin organisms, including S. aureus, coagulase negative staphylococci and Corynebacterium spp. However, P. aeruginosa and Enterobacteriaceae are also frequently isolated, with Candida found as a less frequent pathogen (14). Nienaber et al. evaluated pathogens implicated in local infections (which included DLI, pump pocket infections, cardiovascular implantable electronic device pocket infections, and mediastinitis without bloodstream infection) and found that the most common pathogens isolated were Gram-positive cocci (44.8%) and Gram-negative rods (24.1%) (4).

Polymicrobial infections are also common and may involve multi-drug resistant organisms. In one published case of a progressive polymicrobial infection with methicillin resistant S. aureus (MRSA) and an extended spectrum beta lactamase Erwinia, the patient was eventually treated with pump exchange with no recurrence of infection after 1 year (15). As our understanding of DLI evolution progresses, it is becoming apparent that polymicrobial infections often occur due to superinfection of an existing infected driveline site while a patient remains on suppressive therapy for the initial pathogen. In these cases, P. aeruginosa is one of the most common secondary pathogens (16).

Additionally, it is often difficult to distinguish commensal skin organisms from true pathogens. In a recent example, a Staphylococcus epidermidis DLI was initially suspected, but allergy to an e-polytetrafluoroethylene membrane used during surgery was ultimately identified as the likely cause of the patient’s purulent fluid collections (17).


Pathogenesis

The LVAD driveline provides an ideal surface for the formation of biofilms due to its high surface area, and biofilm formation has been well demonstrated in a murine model of staphylococcal DLI (18). Biofilms are also thought to be involved in the pathogenesis of DLIs due to other biofilm-forming organisms such as P. aeruginosa, Candida spp., and Enterococcus spp.

Biofilms occur when bacteria adhere to a surface and form microcolonies embedded in an extracellular polymeric substance known as matrix. In this environment, the actual organisms may account for as little as 10% of the biomass of the biofilm, with the remainder of the biofilm biomass comprised of matrix. The composition of the matrix varies from species to species and can protect the bacterial cells from the host’s immune system and prevent penetration of antibiotics (19). These factors make antibiotic selection particularly important in the treatment of DLIs.


Treatment

Guidelines

No comprehensive guidelines for treatment of LVADIs exist to date, though general guidelines for treatment duration proposed by Nienaber et al. suggest 2-4 weeks of antimicrobial therapy with or without surgical debridement for DLI. Suppressive antibiotics are not suggested for isolated DLI but should be considered if pump pocket infection is suspected (4). Current guidelines do not address choice of empiric or targeted antimicrobial therapy for LVAD DLIs.

Empiric therapy

For any suspected DLI, a culture of the site should be obtained and empiric therapy started while awaiting results. The best empiric therapy for DLIs has not been established. For early localized infection, oral antibiotics are often used (4). There is no standard definition for early localized infection, but it is generally described as drainage or inflammation around the driveline exit site without systemic symptoms or concern for underlying abscess. The current recommended empiric antibiotic choice for these early infections at the University of Minnesota is doxycycline 100 mg BID for 14 days with antibiotic choice adjusted as needed based on culture results. Another published regimen includes ciprofloxacin 500 mg BID and doxycycline 100 mg PO BID for 10 days (20). For more extensive local infections or when systemic symptoms are present, patients are admitted to the hospital for monitoring, imaging, and broad-spectrum antibiotics while awaiting culture results.

Targeted therapy

When the pathogen(s) responsible for a DLI are identified, empiric antibiotics can be changed to targeted therapy. However, multiple options are often available and antibiotic choice is currently guided by reports of anecdotal success rather than rigorous evidence. For example, several oral antibiotics might be chosen for a superficial MRSA DLI and an equal number of choices exist if intravenous therapy is used. While the potential benefits of certain antibiotic choices can be extrapolated from literature pertaining to other biofilm-related infections, such as osteoarticular infections with retained hardware, significant differences exist between the surfaces involved in those infections and the driveline exit site (21). Despite the growing base of literature around LVADIs, specific guidance for antibiotic choice is currently lacking.

This absence of evidence is particularly perplexing when rifampin is considered as part of a treatment regimen. Rifampin is a bactericidal antibiotic that can penetrate and eradicate biofilms (22). It can be used only as adjunctive therapy with another antibiotic due to potential for development of resistance, but it demonstrates synergy with many common antibiotics making it an appealing addition to therapy for DLIs (23). However, rifampin interacts with warfarin and causes significant INR instability leading to complications such as GIB and CVA (23). To date, neither the bleeding risks nor the potential treatment benefits of rifampin have been studied in patients with LVADIs. However, it is known that infection is a risk factor for GIB in patients with LVADs (24). Additionally, persistent bloodstream infection in patients with LVADs has been associated with all-cause CVA (25). Therefore, defining the risk and benefits of rifampin use in this population is especially important.

Suppressive therapy

Suppressive antimicrobial therapy in LVADIs is often used in cases of extensive or recurrent infection. In fact, Nienaber et al. found that 35% of patients with local infections received suppressive oral antimicrobials (4). However, the risks and benefits of suppression remain unclear. In one oft-cited small study, 5/16 (31%) of patients treated with chronic antimicrobials experienced clinical failure. The study could not draw any firm conclusions due to the small sample size (26). Therefore, evidence to date is insufficient to guide therapy choices.

Device exchange

Device exchange can be used to achieve source control in otherwise intractable DLIs, and 0.6-11% of device exchanges are performed for infection. While one study showed that 6.5% of patients who undergo pump exchange die within 30 days, another showed no difference in mortality between the exchange and non-exchange group (27,28). A recent case series and review of the literature found that data remain limited regarding successful approaches and outcomes for device exchange due to the differences in pathogens, patient characteristics, and approach to management among published cases (29). While device exchange should remain a consideration for difficult infections, larger studies of patients undergoing exchange would help to guide which patients should be offered device exchange, the utility of interventions such as antibiotic impregnated beads, and the ideal use of antibiotics after the procedure.

Similarly, transplant can be used to cure DLIs. In a 2015 study of patients with active DLI at the time of transplant there was no difference found in length of hospital stay, infections, or mortality after 30 months (30). The safety of transplant in patients with poorly controlled infections is unclear, though there are reported cases of good outcomes in these settings (31).

Alternative therapies

The frequent failure of traditional antibiotics to control DLIs has resulted in renewed interest in alternative therapies. Many of the proposed treatment modalities are those currently used in chronic wound management. Some, such as medical grade honey and topical antibiotics are used despite lack of evidence for efficacy in this patient group. Others are currently under investigation. Ultraviolet B (UVB) radiation has been used in chronic wound therapy but not at the driveline exit site due to concerns about effects of UVB on driveline materials. A recent in vitro pilot study suggested that UVB therapy may change the elasticity of the HeartMate II driveline materials but that this change may not be detrimental, which is encouraging as a future avenue of investigation (32). Investigations of UVB as well as other alternative therapies are valuable since these therapies often avoid complications such as antibiotic resistance, medication interactions, and Clostridium difficile infection that can occur with ongoing use of traditional antibiotics.


Prevention

Implantation method and device type

Infection rates may differ between VAD device types, although this has not been shown definitively. A single retrospective observational study of patients receiving an LVAD as BTT by Haglund et al. demonstrated a 17% readmission rate for DLI among HeartMate II recipients vs. a 2% readmission rate for HeartWare HVAD recipients (P=0.03). However, the small sample size (n=81) and the retrospective study design suggest that further investigation is needed to confirm these findings (20). Another recent study showed a DLI rate of 0.25 events per patient-year for the HeartWare HVAD (33). One potential advantage of the HVAD is its smaller driveline diameter since a smaller diameter is thought to be associated with lower infection rates (1,33).

There are ongoing efforts to optimize driveline implantation techniques with the goal of decreasing DLIs (34-36). One promising study by Dean et al. compared DLI rates in patients with the Dacron velour portion of the HeartMate II driveline either partially exposed (control group) or entirely implanted under the skin (silicone-skin interface group). Their results showed 1 year DLI prevalence rates of 9% in the silicone-skin interface group vs. 23% in the control group (37). However, the ideal length of non-Dacron coated silicone to internalize remains unclear (38).

Nutrition

Nutrition has also emerged as a possible predictor of DLI requiring admission. In a study by Imamura et al., serum albumin concentration and low body mass index at hospital discharge after LVAD implantation both independently predicted readmission due to DLI. Based on these findings, the authors developed a scoring system that risk-stratifies patients as low, intermediate, or high risk for DLI requiring admission based on these two parameters. While this study was limited by its design as a single center retrospective study, these findings suggest that optimizing nutrition for patients with recently implanted devices may help to prevent admission due to DLI (39).

Driveline dressings

Typically, the driveline exit site is cleansed with an antiseptic agent such as chlorhexidine, hydrogen peroxide, or povidone-iodine and dressed in a sterile fashion either daily or weekly. The site is then usually stabilized with a driveline securement device. There is currently no standard protocol for changing driveline dressings, although an effort is being made to evaluate the convenience and efficacy of ready-made dressing kits (40). A small 2015 study showed no difference between daily and weekly dressing changes in a prospective trial, but the study was significantly limited because patients were only followed for 30 days after hospital discharge (41).

In another recent study, driveline exit sites were dressed with octenidine with or without the addition of merbromin in a non-blinded study. Over 2 years the merbromin group had no patients with DLI vs. 11.8% of patients in the standard group developing infections (42). This study suggests that we would benefit from closer evaluation of driveline exit site maintenance and alternative dressing strategies as an easy and effective method to reduce or prevent DLIs.


Conclusions

Our understanding of LVAD DLI epidemiology and evolution has improved in recent years. Additionally, initial strides have been taken to optimize LVAD implantation techniques and adjust driveline exit site maintenance as methods for DLI prevention. However, further studies are urgently needed to define evidence-based treatment of these infections including choice of biofilm-active antibiotic regimens, necessity of device exchange, and utility of suppressive antibiotics.


Acknowledgements

I would like to acknowledge Gary M. Dunny for helpful discussions during the preparation of this manuscript. This work was supported by the National Institutes of Health Clinical and Translational Science Award at the University of Minnesota [UL1TR000114].


Footnote

Conflicts of Interest: The author has no conflicts of interest to declare.


References

  1. Pereda D, Conte JV. Left ventricular assist device driveline infections. Cardiol Clin 2011;29:515-27. [PubMed]
  2. Akhter SA, Badami A, Murray M, et al. Hospital Readmissions After Continuous-Flow Left Ventricular Assist Device Implantation: Incidence, Causes, and Cost Analysis. Ann Thorac Surg 2015;100:884-9. [PubMed]
  3. Wickline SA, Fischer KC. Can infections be imaged in implanted devices? ASAIO J 2000;46:S80-1. [PubMed]
  4. Nienaber JJ, Kusne S, Riaz T, et al. Clinical manifestations and management of left ventricular assist device-associated infections. Clin Infect Dis 2013;57:1438-48. [PubMed]
  5. Trachtenberg BH, Cordero-Reyes A, Elias B, et al. A review of infections in patients with left ventricular assist devices: prevention, diagnosis and management. Methodist Debakey Cardiovasc J 2015;11:28-32. [PubMed]
  6. Goldstein DJ, Naftel D, Holman W, et al. Continuous-flow devices and percutaneous site infections: clinical outcomes. J Heart Lung Transplant 2012;31:1151-7. [PubMed]
  7. Koval CE, Rakita R; AST Infectious Diseases Community of Practice. Ventricular assist device related infections and solid organ transplantation. Am J Transplant 2013;13 Suppl 4:348-54. [PubMed]
  8. Hannan MM, Husain S, Mattner F, et al. Working formulation for the standardization of definitions of infections in patients using ventricular assist devices. J Heart Lung Transplant 2011;30:375-84. [PubMed]
  9. Lushaj EB, Badami A, Osaki S, et al. Impact of age on outcomes following continuous-flow left ventricular assist device implantation. Interact Cardiovasc Thorac Surg 2015;20:743-8. [PubMed]
  10. Zierer A, Melby SJ, Voeller RK, et al. Late-onset driveline infections: the Achilles’ heel of prolonged left ventricular assist device support. Ann Thorac Surg 2007;84:515-20. [PubMed]
  11. Levy DT, Minamoto GY, Da Silva R, et al. Role of gallium SPECT-CT in the diagnosis of left ventricular assist device infections. ASAIO J 2015;61:e5-10. [PubMed]
  12. Dell’Aquila AM, Mastrobuoni S, Alles S, et al. Contributory Role of Fluorine 18-Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography in the Diagnosis and Clinical Management of Infections in Patients Supported With a Continuous-Flow Left Ventricular Assist Device. Ann Thorac Surg 2016;101:87-94. [PubMed]
  13. Fujino T, Higo T, Tanoue Y, et al. FDG-PET/CT for driveline infection in a patient with implantable left ventricular assist device. Eur Heart J Cardiovasc Imaging 2016;17:23. [PubMed]
  14. Nienaber J, Wilhelm MP, Sohail MR. Current concepts in the diagnosis and management of left ventricular assist device infections. Expert Rev Anti Infect Ther 2013;11:201-10. [PubMed]
  15. Hamdan R, Ali F, Saab M. Complicated Left Ventricular Assist Device Infection Treated With Pump Exchange and Left Ventricular Apical Resection. Artif Organs 2015;39:981-2. [PubMed]
  16. Koval CE, Thuita L, Moazami N, et al. Evolution and impact of drive-line infection in a large cohort of continuous-flow ventricular assist device recipients. J Heart Lung Transplant 2014;33:1164-72. [PubMed]
  17. Kornberger A, Walter V, Khalil M, et al. Suspected involvement of EPTFE membrane in sterile intrathoracic abscess and pericardial empyema in a multi-allergic LVAD recipient: a case report. J Cardiothorac Surg 2015;10:99. [PubMed]
  18. Toba FA, Akashi H, Arrecubieta C, et al. Role of biofilm in Staphylococcus aureus and Staphylococcus epidermidis ventricular assist device driveline infections. J Thorac Cardiovasc Surg 2011;141:1259-64. [PubMed]
  19. Flemming HC, Wingender J. The biofilm matrix. Nat Rev Microbiol 2010;8:623-33. [PubMed]
  20. Haglund NA, Davis ME, Tricarico NM, et al. Readmissions After Continuous Flow Left Ventricular Assist Device Implantation: Differences Observed Between Two Contemporary Device Types. ASAIO J 2015;61:410-6. [PubMed]
  21. Coiffier G, Albert JD, Arvieux C, et al. Optimizing combination rifampin therapy for staphylococcal osteoarticular infections. Joint Bone Spine 2013;80:11-7. [PubMed]
  22. Tang HJ, Chen CC, Cheng KC, et al. In vitro efficacies and resistance profiles of rifampin-based combination regimens for biofilm-embedded methicillin-resistant Staphylococcus aureus. Antimicrob Agents Chemother 2013;57:5717-20. [PubMed]
  23. Forrest GN, Tamura K. Rifampin combination therapy for nonmycobacterial infections. Clin Microbiol Rev 2010;23:14-34. [PubMed]
  24. Lopilato AC, Doligalski CT, Caldeira C. Incidence and Risk Factor Analysis for Gastrointestinal Bleeding and Pump Thrombosis in Left Ventricular Assist Device Recipients. Artif Organs 2015;39:939-44. [PubMed]
  25. Trachtenberg BH, Cordero-Reyes AM, Aldeiri M, et al. Persistent blood stream infection in patients supported with a continuous-flow left ventricular assist device is associated with an increased risk of cerebrovascular accidents. J Card Fail 2015;21:119-25. [PubMed]
  26. Jennings DL, Chopra A, Chambers R, et al. Clinical outcomes associated with chronic antimicrobial suppression therapy in patients with continuous-flow left ventricular assist devices. Artif Organs 2014;38:875-9. [PubMed]
  27. Moazami N, Milano CA, John R, et al. Pump replacement for left ventricular assist device failure can be done safely and is associated with low mortality. Ann Thorac Surg 2013;95:500-5. [PubMed]
  28. Anand J, Singh SK, Hernández R, et al. Continuous-flow ventricular assist device exchange is safe and effective in prolonging support time in patients with end-stage heart failure. J Thorac Cardiovasc Surg 2015;149:267-75, 278.e1.
  29. Levy DT, Guo Y, Simkins J, et al. Left ventricular assist device exchange for persistent infection: a case series and review of the literature. Transpl Infect Dis 2014;16:453-60. [PubMed]
  30. Bhatia N, Voelkel AJ, Hussain Z, et al. Safety and Feasibility of Induction Immunosuppression When Driveline Infection Is an Indication for Cardiac Transplantation. Thorac Cardiovasc Surg 2015;63:675-83. [PubMed]
  31. Swartz TH, Huprikar S, Labombardi V, et al. Heart transplantation in a patient with heteroresistant vancomycin-intermediate Staphylococcus aureus ventricular assist device mediastinitis and bacteremia. Transpl Infect Dis 2013;15:E177-81. [PubMed]
  32. Evans AC, Wright GA, McCandless SP, et al. Ultraviolet Radiation Affects Thoratec HeartMate II Driveline Mechanical Properties: A Pilot Experiment. ASAIO J 2015;61:731-3. [PubMed]
  33. John R, Aaronson KD, Pae WE, et al. Drive-line infections and sepsis in patients receiving the HVAD system as a left ventricular assist device. J Heart Lung Transplant 2014;33:1066-73. [PubMed]
  34. Fleissner F, Avsar M, Malehsa D, et al. Reduction of driveline infections through doubled driveline tunneling of left ventricular assist devices. Artif Organs 2013;37:102-7. [PubMed]
  35. Entwistle J. Intraoperative positioning to prevent driveline insulation from acting as a conduit for left ventricular assist device pocket infection. J Thorac Cardiovasc Surg 2014;148:2437. [PubMed]
  36. Pinninti M, Thohan V, Sulemanjee NZ. Driveline insulation as a conduit for left ventricular assist device pocket infection. J Thorac Cardiovasc Surg 2014;148:e135-6. [PubMed]
  37. Dean D, Kallel F, Ewald GA, et al. Reduction in driveline infection rates: Results from the HeartMate II Multicenter Driveline Silicone Skin Interface (SSI) Registry. J Heart Lung Transplant 2015;34:781-9. [PubMed]
  38. Rubinfeld G, Levine JP, Reyentovich A, et al. Management of Rapidly Ascending Driveline Tunnel Infection. J Card Surg 2015;30:853-5. [PubMed]
  39. Imamura T, Kinugawa K, Nitta D, et al. Readmission due to driveline infection can be predicted by new score by using serum albumin and body mass index during long-term left ventricular assist device support. J Artif Organs 2015;18:120-7. [PubMed]
  40. Iseler J, Hadzic KG. Developing a kit and video to standardize changes of left ventricular assist device dressings. Prog Transplant 2015;25:224-9. [PubMed]
  41. Wus L, Manning M, Entwistle JW 3rd. Left ventricular assist device driveline infection and the frequency of dressing change in hospitalized patients. Heart Lung 2015;44:225-9. [PubMed]
  42. Menon AK, Baranski SK, Unterkofler J, et al. Special Treatment and Wound Care of the Driveline Exit Site after Left Ventricular Assist Device Implantation. Thorac Cardiovasc Surg 2015;63:670-4. [PubMed]
Cite this article as: Leuck AM. Left ventricular assist device driveline infections: recent advances and future goals. J Thorac Dis 2015;7(12):2151-2157. doi: 10.3978/j.issn.2072-1439.2015.11.06

Download Citation